University of Kirikkale Medical School

z-ACT (Celite Activator) from ATS Biomedical, Turkey tubes were tested with different ACT-Trol control solutions

specified by using Hemochron 401 equipment in Department of Cardiovascular Surgery and

Central Laboratory, University of Kirikkale, Turkey.

Results are documented below :

  Level I (LOT # 1-61-8) Level II (LOT # 1-45-17) Level III (LOT # 1-61-9)
1 121 221 336
2 130 237 314
3 123 237 321
4 131 235 317
5 135 245 328
6 126 221 341
7 126 251 327
8 127 243 320
9 121 230 333
10 124 223 319
11 130 226 336
12 127 238 320
13 116 235 329
14 136 237 331
15 128 223 319
16 126 225 327
17 119 226 329
18 127 237 319
19 137 234 328
20 134 228 337

Should you have any inquiries, please do not hesitate to contact us on the following address.

 

S. Gunaydin, MD, PhD 

Chair, Department of Cardiovascular Surgery

University of Kirikkale, Turkey

 

 